Related Experiment Video
Updated: Jul 7, 2026

Assessment of Physical Activity Intensity with Accelerometers and Oxygen Consumption
Published on: June 20, 2025
Concurrent validity of the PAM accelerometer relative to the MTI Actigraph using oxygen consumption as a reference
S M Slootmaker1, M J M Chin A Paw, A J Schuit
1EMGO Institute, Department of Public and Occupational Health, and Body@Work Research Center Physical Activity, Work and Health, TNO-VU University Medical Center, Amsterdam, The Netherlands.
The Personal Activity Monitor (PAM) accelerometer shows high reliability and comparable validity to Actigraph for measuring physical activity. This device can accurately rank individuals by energy expenditure, aiding in habitual physical activity monitoring.
Area of Science:
- Physical Activity Measurement
- Biomedical Engineering
- Exercise Physiology
Background:
- Objective physical activity monitoring is crucial for public health.
- Accelerometers are widely used to assess physical activity and energy expenditure.
- Validating new devices against established measures is essential.
Purpose of the Study:
- To determine the concurrent validity of the Personal Activity Monitor (PAM) against the Actigraph accelerometer, using oxygen consumption (VO2) as a reference.
- To evaluate the test-retest reliability of the PAM device.
Main Methods:
- Thirty-two healthy adults performed treadmill and stair walking wearing PAM, Actigraph, and Cosmed K4b2.
- Concurrent validity was assessed using correlation coefficients and agreement in energy expenditure (EE).
- Test-retest reliability of 296 PAM devices was evaluated using a laboratory shaker, calculating intraclass correlation coefficients (ICC) and coefficient of variation (CV).
Main Results:
- PAM and Actigraph showed strong correlations with VO2 during treadmill walking (r(2)=0.82 and 0.64, respectively) and stair walking (r(2)=0.93 and 0.74, respectively).
- PAM and Actigraph accelerometers demonstrated comparable validity (r(2)=0.95 for treadmill, r(2)=0.65 for stair walking).
- PAM exhibited high test-retest reliability (ICC=0.80, intra-CV=1.5%). Both devices underestimated EE.
Conclusions:
- The Personal Activity Monitor (PAM) is a valid and reliable accelerometer for assessing physical activity.
- PAM demonstrates comparable validity to Actigraph and can effectively rank individuals by energy expenditure.
- PAM is suitable for objective monitoring of habitual physical activity.