Triangle ventricular pacing in a non-responder to conventional bi-ventricular pacing
Kentaro Yoshida1, Yasuhiro Yokoyama, Yoshihiro Seo
1Cardiovascular Division, Institute of Clinical Medicine, University of Tsukuba, 1-1-1 Tennodai, Tsukuba 305-8575, Ibaraki, Japan. kentaroyo@nifty.com
Abstract:
A 56-year-old man with cardiomyopathy secondary to myotonic dystrophy and complete atrioventricular block presented to our institution. A cardiac resynchronization therapy (CRT) device (InSync 8040, Medtronic, Inc., Minneapolis, MN, USA) was implanted by a conventional bi-ventricular pacing (Bi-V) technique. However, the patient's NYHA class did not improve from class IV despite optimized medical therapy. One month after the CRT device implantation, we altered the pacing configuration from that of Bi-V to 'triangle ventricular pacing' (Tri-V), i.e. conventional Bi-V from the right ventricular (RV) apex and left ventricle plus additional pacing from the RV outflow tract, using a Y connector to bifurcate the RV bipolar output of the device into an anode and a cathode. In both the acute and 3 month follow-up studies, objective parameters revealed better resynchronization effects with Tri-V, and the patient's NYHA class immediately improved to class II. Triangle ventricular pacing may have the potential to decrease the number of non-responders to CRT.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...