Ring Char Formation During High-Power, Short-Duration Ablation Using an Irrigated Six-Hole Contact Force Sensing Tip
Yasuteru Yamauchi1, Yuichiro Sagawa1, Kazuya Murata1
1Department of Cardiology Japan Red Cross Yokohama City Bay Hospital Kanagawa Japan.
Journal of Arrhythmia
|February 18, 2026
Summary
High-power, short-duration radiofrequency ablation for pulmonary veins can cause thrombus formation. This complication occurred during ablation of the right pulmonary veins, leading to an impedance rise and visible catheter tip thrombus.
Area of Science:
- Electrophysiology
- Cardiovascular Medicine
- Medical Device Technology
Background:
- Circumferential ablation of pulmonary veins (PVs) is a key procedure in treating atrial fibrillation.
- High-power, short-duration (HPSD) ablation is an advanced technique aiming for efficient lesion creation.
- Potential complications during radiofrequency (RF) ablation, such as impedance changes, require careful monitoring.
Purpose of the Study:
- To report a specific complication encountered during HPSD ablation of the right PVs.
- To describe the characteristics of thrombus formation associated with HPSD ablation.
- To highlight the importance of recognizing impedance changes as a potential indicator of thrombus.
Main Methods:
- Performed circumferential ablation of the right PVs using HPSD energy.
- Monitored impedance changes during RF delivery.
- Visualized catheter tip and surrounding structures after an observed impedance increase.
Main Results:
- An abrupt increase in impedance from 120 to 180 Ω occurred 11 seconds after initiating RF delivery.
- Withdrawal of the ablation catheter revealed a circumferential thrombus adhered to the catheter tip, above the irrigation holes.
- The removed thrombus was ring-shaped, confirming adherence to the catheter tip post-ablation.
Conclusions:
- HPSD ablation, particularly at the anterior wall of the right PVs, carries a risk of catheter tip thrombus formation.
- Abrupt impedance increases during RF ablation can signify the development of thrombus.
- Awareness of this complication is crucial for procedural safety and patient outcomes in PV ablation.