Related Experiment Video
Updated: Jul 7, 2026

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Optimizing color reproduction of a topometric measurement system for medical applications
Matthias Westhäuser1, Guido Bischoff, Zoltan Böröcz
1Laboratory of Biophysics, University of Muenster, D-48129 Muenster, Germany.
Abstract:
Integration of color information in 3D models acquired with a topometric measurement system opens a wide field of applications in medicine, e.g. prosthetics. To ensure an accurate and reliable color reproduction in a 3D topometric measurement system, different approaches have been evaluated. These techniques focus on the calibration of the CCD camera using color test charts by applying a polynomial regression method as well as numerical methods to compensate an intensity decay of the illumination in 3D datasets which is caused by varying distances and orientations of the measured object with respect to the light source. The methods presented in this contribution have been applied to a topometric measurement system with one-chip color CCD cameras. An enhanced color reproduction of various objects is demonstrated.

