Related Experiment Video
Updated: Jul 7, 2026

Transcorporal Artificial Urinary Sphincter Cuff Placement in a Case Requiring Revision for Urethral Atrophy
Published on: June 16, 2022
Recurrent stress urinary incontinence after dislodged screws in patient with bone-anchored suburethral sling
Ayman Mahdy1, Mostafa Elmissiry, Gamal Ghoniem
1Section of Voiding Dysfunction, Female Urology, and Reconstructive Surgery, Cleveland Clinic Florida, Weston, Florida 33331, USA.
Abstract:
This is the first reported case of recurrent stress urinary incontinence secondary to dislodged bone-anchoring screws. A 71-year-old man who had undergone successful bone-anchored sling placement for post-prostatectomy incontinence, presented with recurrent stress urinary incontinence after a bout of coughing. Pelvic radiography revealed three screws had been dislodged. The mesh was left in place, with replacement of the dislodged screws and the addition of one extra screw. The findings of this case suggest that dislodged screws can be a possible cause of sling failure. Preoperative bone scanning and postoperative pelvic radiography should be a part of the routine workup for patients with bone-anchored slings.