Related Experiment Video
Updated: Jul 7, 2026

Angle-resolved Photoemission Spectroscopy At Ultra-low Temperatures
Published on: October 9, 2012
Development of a vacuum ultraviolet laser-based angle-resolved photoemission system with a superhigh energy
Guodong Liu1, Guiling Wang, Yong Zhu
1National Laboratory for Superconductivity, Beijing National Laboratory for Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100080, People's Republic of China.
Abstract:
The design and performance of the first vacuum ultraviolet (VUV) laser-based angle-resolved photoemission (ARPES) system are described. The VUV laser with a photon energy of 6.994 eV and bandwidth of 0.26 meV is achieved from the second harmonic generation using a novel nonlinear optical crystal KBe2BO3F2. The new VUV laser-based ARPES system exhibits superior performance, including superhigh energy resolution better than 1 meV, high momentum resolution, superhigh photon flux, and much enhanced bulk sensitivity, which are demonstrated from measurements on a typical Bi2Sr2CaCu2O8 high temperature superconductor. Issues and further development related to the VUV laser-based photoemission technique are discussed.
Related Concept Videos
Transmission Electron Microscopy
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Overview of Electron Microscopy
Super-resolution Fluorescence Microscopy