Assembly, Tuning and Use of an Apertureless Near Field Infrared Microscope for Protein Imaging
Published on: November 25, 2009
Balanced homodyning for apertureless near-field optical imaging
M Labardi1, O Tikhomirov, C Ascoli
1INFM-CNR polyLab, Dipartimento di Fisica Enrico Fermi, Università di Pisa, Largo B. Pontecorvo 3, I-56127 Pisa, Italy. labardi@df.unipi.it
Abstract:
A quadrature optical detection technique, based on polarized balanced-homodyne interferometry, has been developed for specific application to apertureless near-field scanning optical microscopy (ANSOM). With such technique, multiplicative background interference, inficiating quantitative optical imaging in standard homodyne-based ANSOM, can be suppressed. Periodic modulation of interferometric optical phase, typically employed in heterodyne-based ANSOMs even to such purpose, is not needed in the present configuration. Homodyne detection also facilitates detection of harmonic components of the ANSOM optical signal at the probe/sample distance modulation frequency, necessary for near-field discrimination and suppression of artifacts. Furthermore, since amplitude signal is not affected by phase fluctuations generated in the optical path of the interferometer, an optical fiber could be included in one interferometer arm, to couple the ANSOM head to the detection system, obtaining improved versatility of the instrument. A demonstration of the interferometer performance is given by a test confocal optical scan of a mirror surface. This technique, as applied to near-field microscopy, is anticipated to provide absolute values of optical contrast not depending on background interference and topography artifacts.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...