Fire/burn risk with electrosurgical devices and endoscopy fiberoptic cables
1Department of Otolaryngology, Leonard M. Miller School of Medicine, University of Miami, Miami, Florida 33136, USA.
Purpose:
The purpose of the study was to systematically explore the fire and burn risk associated with fiberoptic cables and electrosurgical devices.
Materials And Methods:
A 300-W light source was connected to a standard gray fiberoptic light cable. The end of the cable was either rested atop or buried within a cotton towel or polypropylene drape in the presence or absence of 100% oxygen for up to 10 minutes. A monopolar electrosurgical device set at 1 W, 10 W, or 30 W was tested on a cotton towel or polypropylene drape for a period of 30 seconds. All trials were repeated.
Results:
Resting the light cable on top of the cotton towel or polypropylene drape with or without oxygen produced no result. Burying the end of the cable within the drape produced a hole in the drape within 15 seconds both with and without oxygen. Burying the end of the cable within the cotton towel produced a yellow discoloration after 2 minutes both with and without oxygen. The monopolar electrosurgical device set at 30 W burned immediately through the polypropylene drape, producing a skin burn. All other trials with monopolar electrocautery produced no result. No flame or fire was produced in any trial.
Conclusions:
Fiberoptic cables and electrosurgical generators represent a serious burn risk for surgical patients, with operating room drapes and towels affording only limited protection. Otolaryngologists should be keenly aware of the risks that these devices represent because our specialty uses them frequently.
More Related Videos
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Burn Injuries
The damage results in the death of skin cells, which can lead to a massive loss of fluid. Dehydration, electrolyte imbalance, and renal and circulatory failure follow, which can be fatal. Burn patients are treated with intravenous fluids to offset...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):