QA programme for radon and its short-lived progeny measuring instruments in NRPI Prague
1National Radiation Protection Institute, Department of Radon Expert Group, Prague, Czech Republic. karel.jilek@suro.cz
Abstract:
To subserve the institutional research and tasks coming out from the Czech National Radon Programme, a new QA programme to calibrate all the known types of devices that measure radon and its short-lived progeny was developed at the Department of Radon mobile group of the National Radiation Protection Institute (NRPI) at Prague. The programme also included calibration of instruments measuring a unique quantity of unattached and attached fractions of short- lived radon progeny Generally, NRPI declares estimation of radon concentration during all routine calibration measurements with an overall uncertainty <5% (one sigma) and of equilibrium-equivalent radon concentration with an overall uncertainty <10% (one sigma). The results of the comparative measurements of the unattached and attached fractions of each short-lived radon progeny carried out with a comparing continuous monitor Fritra 4 in the German reference radon chamber at PTB Braunschweig indicated an acceptable level of agreement, up to 10%.
Related Concept Videos
Radioactive Decay and Radiometric Dating
Radioactivity and Nuclear Equations
A nuclide of an element has a specific number of protons and...

![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)