Regional difference in compression artifacts in low-dose chest CT images: effects of mathematical and perceptual
Kil Joong Kim1, Bohyoung Kim, Kyoung Ho Lee
1Department of Radiology, Seoul National University Bundang Hospital, Seoul National University College of Medicine, Institute of Radiation Medicine, and Seoul National University Medical Research Center, Gyeonggi-do, Seoul 463-707, Korea.
Objective:
The objective of our study was to investigate the difference of perceptible artifacts between the lungs and the chest wall and mediastinum in Joint Photographic Experts Group (JPEG) 2000-compressed low-dose chest CT images and to show that a perceptual image quality metric-the High-Dynamic Range Visual Difference Predictor (HDR-VDP)-can reproduce this regional difference.
Materials And Methods:
Twenty images were compressed reversibly and irreversibly to 6:1-30:1. To analyze the two regions separately (lungs; and chest wall and mediastinum), the compressed pixels outside each tested region were replaced with the original pixels. By comparing the compressed and original images, three radiologists independently rated the compression artifacts as grade 0, none, indistinguishable; 1, barely perceptible; 2, subtle; or 3, significant. At each compression level, the two regions were compared for the readers' responses, peak signal-to-noise ratio (PSNR), and HDR-VDP results. Wilcoxon's signed rank tests and exact tests for paired proportions were used with a p value threshold of 0.05.
Results:
Artifacts were rated as lower grades in the lungs than in the chest wall and mediastinum, showing statistical significances at 10:1-20:1 for reader 1, 8:1-15:1 for reader 2, and 8:1-20:1 for reader 3. Grade 0 was more frequent in the lungs, showing statistical significances at 10:1 for reader 1 and at 8:1-10:1 for readers 2 and 3. The results of PSNR indicated greater artifacts in the lungs (p < 0.001), whereas HDR-VDP results indicated fewer artifacts in the lungs (p < 0.001).
Conclusion:
Although compression artifacts are mathematically greater in the lungs than in the chest wall and mediastinum, radiologists' artifact perceptions are the opposite, which can be successfully reproduced by HDR-VDP.
More Related Videos
08:05Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
Published on: December 19, 2020
08:02Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
Published on: November 15, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT