Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Investigation of the relationship between linear attenuation coefficients and CT Hounsfield units using radionuclides
Saxby Brown1, Dale L Bailey, Kathy Willowson
1Institute of Medical Physics, School of Physics, University of Sydney, NSW 2006, Australia. sbrown@physics.usyd.edu.au
Abstract:
This study has investigated the relationship between linear attenuation coefficients (mu) and Hounsfield units (HUs) for six materials covering the range of values found clinically. Narrow-beam mu values were measured by performing radionuclide transmission scans using (99m)Tc, (123)I, (131)I, (201)Tl and (111)In. The mu values were compared to published data. The relationships between mu and HU were determined. These relationships can be used to convert computed tomography (CT) images to mu-maps for single photon emission computed tomography (SPECT) attenuation correction.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...