Improved sectioning in a slit scanning confocal microscope
Related Experiment Video
Updated: Jul 2, 2026

Video-rate Scanning Confocal Microscopy and Microendoscopy
Published on: October 20, 2011
Vincent Poher1, Gordon T Kennedy, Hugh B Manning
1Photonics Group, Blackett Laboratory, Department of Chemistry, Imperial College London, London, UK. vincent.poher03@imperial.ac.uk
Abstract:
We describe a simple implementation of a slit scanning confocal microscope to obtain an axial resolution better than that of a point-scanning confocal microscope. Under slit illumination, images of a fluorescent object are captured using an array detector instead of a line detector so that out-of-focus light is recorded and then subtracted from the adjacent images. Axial resolution after background subtraction is 2.2 times better than the slit confocal resolution, and out-of-focus image suppression is calculated to attenuate with defocus faster by 1 order of magnitude than in the point confocal case.