Safety and usefulness of non-occlusion image acquisition technique for optical coherence tomography
Hideaki Kataiwa1, Atsushi Tanaka, Hironori Kitabata
1Department of Cardiovascular Medicine, Wakayama Medical University, Wakayama, Japan. hkataiwa@wakayama-med.ac.jp
Background:
Because optical coherence tomography (OCT) requires balloon-occlusion to overcome image attenuation by blood, the present study investigated the safety and usefulness of a new non-occlusive technique.
Methods And Results:
The 40 angina patients were assigned to the continuous-flushing method or the balloon-occlusion method group. The continuous-flushing method was superior for observing proximal lesions compared with the balloon-occlusion method (94% vs 55%, p=0.01). There were no differences between groups in the visible length and image quality. There were no major complications during OCT.
Conclusions:
The continuous-flushing method is safe and useful for OCT.