Related Experiment Video
Updated: Jul 2, 2026

Transcorporal Artificial Urinary Sphincter Cuff Placement in a Case Requiring Revision for Urethral Atrophy
Published on: June 16, 2022
Cylinder aneurysm of parylene-coated American Medical System (AMS) 700CX penile prosthesis
Joel M Kaufman1, Thomas E Weldon
1Urology Research Options, Aurora, CO 80012, USA. joelk1123@comcast.net
Introduction:
Cylinder aneurysms and leakages are uncommon with three-layered American Medical System (AMS) CX cylinders. Since 2001, an additional parylene coating improves cylinder wear.
Aim:
To report two patients in whom major cylinder aneurysms developed less than 4 years after implantation.
Methods:
Two patients in separate urologic private practices developed significant cylinder aneurysms requiring reoperation.
Results:
Both patients developed aneurysms in 21-cm parylene-coated CX cylinders just short of 4 years of inflatable penile prosthesis (IPP) placement. Both did well after the explantation and insertion of Coloplast Titan IPP (Coloplast; Minneapolis, MN, USA).
Conclusion:
Longer AMS CX IPP cylinders may develop aneurysms more than 3 years after insertion.