Validation of creatinine assays utilizing HPLC and IDMS traceable standards in sera of children
George J Schwartz1, Tai Kwong, Brian Erway
1University of Rochester Medical Center, Rochester, NY, USA. George_Schwartz@urmc.rochester.edu
Abstract:
The purpose of this study was to validate serum creatinine (SCr) concentrations assayed in the Central Biochemistry Laboratory of the National Institutes of Health (NIH)-funded Chronic Kidney Disease in Children (CKiD) study utilizing an enzymatic assay (Siemens Advia 2400) against a method traceable to reference isotope dilution mass spectroscopy (IDMS) developed by the National Institute of Standards and Technology (NIST). High-performance liquid chromatography (HPLC) measured SCr after external validation utilizing IDMS-based standard reference materials. Sera from the first 201 subjects enrolled in CKiD were analyzed and compared for creatinine concentration by enzymatic and HPLC methods. Fifty "normal" pediatric sera were subsequently analyzed. Finally, a "pediatric" reference standard was prepared and examined for accuracy and precision. Enzymatic SCr concentrations (median 1.4 mg/dl) of CKiD subjects were well correlated with HPLC (r = 0.984) but were slightly higher (+7%; p < 0.001). Agreement was poorer at lower SCr (median 0.4 mg/dl) when using samples from normal children and the "pediatric" reference standard. However, the Roche enzymatic assay was comparable with HPLC in accuracy and precision. Referring physicians should be aware of the accuracy and reproducibility of their laboratory's SCr assay. Our enzymatic assay agreed well with HPLC in CKiD subjects with elevated SCr. We suggest that NIST develop a pediatric SCr standard reference material for use by assay manufacturers to improve accuracy and precision of assays at the low SCr levels observed in most pediatric patients.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Serum Studies: Renal Function Tests
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Data Validation
Key parameters for method validation include: