Blood pressure estimation in the human fetal descending aorta
P C Struijk1, V J Mathews, T Loupas
1Department of Obstetrics and Gynecology, Division of Obstetrics and Prenatal Medicine, Erasmus Medical Center, Rotterdam, The Netherlands. p.struijk@erasmusmc.nl
Objectives:
The objectives of this study were to estimate fetal blood pressure non-invasively from two-dimensional color Doppler-derived aortic blood flow and diameter waveforms, and to compare the results with invasively derived human fetal blood pressures available from the literature.
Methods:
Aortic pressures were calculated from digitally recorded color Doppler cineloops of the fetal descending aorta by applying the Womersley model in combination with the two-element Windkessel model, assuming constant pulse wave velocity during the second half of pregnancy. The results were compared with invasively derived human fetal blood pressures obtained from the literature.
Results:
In 21 normal pregnancies the estimated mean aortic pressure regression line increased linearly from 28 mmHg at 20 weeks of gestation to 45 mmHg at 40 weeks of gestation. The pulse pressure based on the regression line increased linearly from 21 mmHg at 20 weeks of gestation to 29 mmHg at 40 weeks of gestation. The aortic compliance exhibited a log linear relationship with the gestational age and a statistically significant eightfold increase was observed between 20 and 40 weeks. The aortic downstream peripheral resistance exhibited an exponentially decaying relationship across the same gestational age range. Non-invasively derived aortic systolic and diastolic aortic pressures were comparable with previously reported invasively derived systolic and diastolic umbilical arterial pressures; however, the mean pressures differed significantly from those reported in the umbilical artery in a separate study. The aortic systolic pressures calculated in this study were significantly higher than invasively derived left ventricular systolic pressures that have been previously reported in the literature.
Conclusions:
This study demonstrates the feasibility of estimating arterial blood pressure in the human fetus. The method described is of potential use in assessing fetal blood pressure non-invasively, particularly for studying relative changes with time.
Related Concept Videos
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the diastolic...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure