Related Experiment Video
Updated: Jun 28, 2026

Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
Evaluation of the Hitachi 717 analyser
1Servei d'Anàlisis Clíniques Hospital Germans Trías i Pujol Carretera de Canyet s/n Badalona Barcelona 08916 Spain.
This study evaluated the Hitachi 717 analyzer under both controlled and routine conditions. The device showed acceptable photometric accuracy at 340 and 405 nm wavelengths. Imprecision values were within expected ranges for most tests. However, linearity dispersion was noted at low absorbance for NADH at 340 nm. The pipette delivery system had increased imprecision at 3 microliters, but reagent pipette performance was acceptable. Routine testing of seven biochemical assays revealed within-run and between-run imprecision that met clinical standards. Some contamination was observed, but relative inaccuracy remained good. These findings suggest the Hitachi 717 is suitable for clinical use within the tested parameters.
Area of Science:
- Clinical laboratory instrumentation
- Analytical chemistry in diagnostics
- Medical device validation
Background:
Clinical laboratories rely on accurate and reliable analyzers to ensure diagnostic precision. Prior research has shown that photometric accuracy and precision are critical for reliable biochemical assays. However, no prior work had resolved how specific wavelengths and sample volumes affect performance in real-world settings. This gap motivated a detailed evaluation of the Hitachi 717 analyzer. The Comisión de Instrumentación de la Sociedad Española de Química Clinica and the European Committee for Clinical Laboratory Standards provide frameworks for such assessments. These guidelines emphasize photometric accuracy, linearity, and imprecision metrics. Yet, few studies have combined these standards with routine operational data. This paper's contribution is a dual-phase evaluation focusing on both analytical unit performance and real-world use. The study distinguishes between laboratory-controlled conditions and typical clinical workflows. By addressing these aspects, the research aims to clarify the device's suitability for clinical use.
Purpose Of The Study:
The study aimed to assess the Hitachi 717 analyzer's performance under both controlled and routine conditions. This evaluation was guided by established clinical laboratory standards. The researchers sought to measure photometric accuracy, imprecision, and linearity across specific wavelengths. They also examined the pipette delivery system and temperature control. Routine testing focused on seven biochemical determinations. The goal was to determine if the analyzer meets clinical requirements for precision and accuracy. The study's motivation stems from the need for reliable diagnostic tools in clinical settings. By evaluating both technical and operational aspects, the research provides a comprehensive performance profile.
Main Methods:
The evaluation followed a two-step process. First, the analytical units were tested for photometric accuracy at 340 and 405 nm. Second, routine operations were assessed using seven biochemical assays. Photometric inaccuracy was measured at these wavelengths. Linearity was evaluated by analyzing absorbance dispersion for NADH at 340 nm. The pipette delivery system was tested at 3 microliters. Reagent pipette performance was also measured. Temperature control was assessed for consistency. Routine testing included glucose, creatinine, iron, total protein, AST, ALP, and calcium. Imprecision was calculated as coefficient of variation (CV) for within-run and between-run measurements.
Main Results:
The photometric inaccuracy was acceptable at 340 and 405 nm. Imprecision at 340 nm ranged from 0.12 to 0.95%. At 405 nm, it ranged from 0.30 to 0.73%. Linearity showed some dispersion at low absorbance for NADH at 340 nm. Drift was negligible. The sample pipette imprecision increased at 3 microliters. Reagent pipette performance was acceptable. Temperature control was good. Routine testing showed within-run CV from 0.6% for AST to 6.9% for iron. Between-run CV ranged from 2.4% for glucose to 9.7% for iron. Carry-over contamination was observed in some cases. Relative inaccuracy remained good across all assays.
Conclusions:
The Hitachi 717 analyzer demonstrated acceptable photometric accuracy at 340 and 405 nm. Imprecision values were within acceptable ranges for most wavelengths. Linearity dispersion at low absorbance for NADH suggests potential limitations. The pipette delivery system showed increased imprecision at 3 microliters. Reagent pipette performance and temperature control met expectations. Routine testing revealed within-run and between-run imprecision that aligns with clinical standards. Carry-over contamination was noted but did not compromise overall accuracy. Relative inaccuracy remained good for all tested constituents. These findings suggest the device is suitable for clinical use within the tested parameters.
Frequently Asked Questions
The photometric inaccuracy was acceptable at 340 and 405 nm wavelengths. Imprecision ranged from 0.12 to 0.95% at 340 nm and 0.30 to 0.73% at 405 nm.
The sample pipette imprecision increased at 3 microliters, while the reagent pipette performance remained acceptable.
Linearity showed some dispersion at low absorbance for NADH at 340 nm, indicating potential limitations in this range.
Within-run imprecision for iron was 6.9%, and between-run imprecision was 9.7%.
Some contamination was observed in the carry-over study, but relative inaccuracy remained good across all assays.
The study suggests the device is suitable for clinical use within the tested parameters, with acceptable accuracy and precision.
Related Concept Videos
Mass Analyzers: Overview
Mass Analyzers: Common Types
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Development of Analytical Methods
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...

