Related Experiment Video
Updated: Jun 28, 2026

Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Investigation of a MOSFET dosimetry system for midpoint dose verification in prostate 3D CRT/IMRT
1School of Chemistry and Physics, Department of Physics, University of Adelaide, Adelaide, Australia. timothy.wiese@adelaide.edu.au
Abstract:
The suitability of MOSFETs (Metal Oxide Semiconductor Field Effect Transistors) for use in in-vivo dosimetry for IMRT prostate treatment and patient setup errors has been investigated in this work. MOSFETs were placed on entrance and exit surfaces of a number of different phantoms (with varying complexities from homogeneous to anthropomorphic). Dose measurements were then used to calculate a midpoint dose, which was compared with an IC placed at the isocentre. The agreements found between the calculated (MOSFETs) and the measured midpoint dose (IC) was: 0.7% for a prostate treatment verification and 3.5% for an IMRT treatment. MOSFETs placed on entry and exit surfaces can detect patient setup offsets of 2 cm, but do not have the sensitivity to confidently detect offsets of 1 cm or smaller.