Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
Mean free path and Mean free time01:22

Mean free path and Mean free time

Consider the gas molecules in a cylinder. They move in a random motion as they collide with each other and change speed and direction. The average of all the path lengths between collisions is known as the "mean free path."

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Optical creation of a supercrystal with three-dimensional nanoscale periodicity.

Nature materials·2019
Same author

Pump-probe spectrometer for measuring x-ray induced strain.

Optics letters·2016
Same author

Transient crystalline superlattice generated by a photoacoustic transducer.

Structural dynamics (Melville, N.Y.)·2016
Same author

X-ray photonic microsystems for the manipulation of synchrotron light.

Nature communications·2015
Same author

Bunch mode specific rate corrections for PILATUS3 detectors.

Journal of synchrotron radiation·2015
Same author

Picosecond time-resolved laser pump/X-ray probe experiments using a gated single-photon-counting area detector.

Journal of synchrotron radiation·2009

Related Experiment Video

Updated: Jun 28, 2026

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers
09:49

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers

Published on: October 23, 2018

Empirical dead-time corrections for synchrotron sources.

D A Walko1, D A Arms, E C Landahl

  • 1Advanced Photon Source, Argonne National Laboratory, Argonne, IL 60439, USA. d-walko@anl.gov

Journal of Synchrotron Radiation
|October 29, 2008
PubMed
Summary

This study compares dead-time correction models for photon-counting detectors at synchrotron light sources. Detector speed and synchrotron fill patterns are critical for accurate dead-time correction.

More Related Videos

X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
10:16

X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells

Published on: August 20, 2019

Related Experiment Videos

Last Updated: Jun 28, 2026

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers
09:49

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers

Published on: October 23, 2018

X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
10:16

X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells

Published on: August 20, 2019

Area of Science:

  • Experimental physics
  • Detector science
  • Synchrotron radiation

Background:

  • Photon-counting detectors are essential for X-ray measurements at synchrotron facilities.
  • Accurate dead-time correction is crucial for reliable data acquisition.
  • Synchrotron sources, like the Advanced Photon Source, utilize various operating modes, including asymmetric fill patterns.

Purpose of the Study:

  • To experimentally compare different dead-time correction models for photon-counting detectors.
  • To evaluate detector performance under various synchrotron operating conditions, especially asymmetric fill patterns.
  • To identify the key factors influencing the selection of appropriate dead-time correction methods.

Main Methods:

  • Systematic performance evaluation of multiple photon-counting detectors.
  • Testing detectors across three operating modes of the Advanced Photon Source.
  • Focus on performance under asymmetric synchrotron fill patterns.
  • Evaluation of several established dead-time correction formulas.

Main Results:

  • Detector speed significantly impacts the accuracy of dead-time corrections.
  • Synchrotron fill patterns, particularly asymmetric ones, critically affect detector performance.
  • The choice of dead-time correction model must account for both detector characteristics and beam conditions.
  • Simple correction formulas show varying degrees of effectiveness depending on the scenario.

Conclusions:

  • The selection of a dead-time correction model is highly dependent on detector speed and the specific synchrotron fill pattern.
  • Understanding these parameters is vital for optimizing data quality in high-flux experiments.
  • Further research may be needed to develop more robust correction models for complex synchrotron conditions.