Utilization of standards generated in the process of thermal decomposition chemically modified silica gel for a single point calibration of a GC/FID system
Related Experiment Video
Updated: Jun 28, 2026

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
M Prokopowicz1, E Luboch, J Namieśnik
1Chemistry Faculty, Technical University of Gdańsk, 80-952 Gdańsk, Poland.
Abstract:
The utilization of a multicomponent gaseous standard mixture, containing CO and CO(2) and obtained by thermal decomposition of a so-called immobilized compound, for a single point calibration of a GC/FID system are described. The generation of such a mixture takes place as a result of thermal decomposition of a sample of chemically modified silica gel placed in a heated chamber of thermal desorber coupled with the device being calibrated via a catalytic methanizer. The mean amount of the analyte librated from unit mass of the gel was 0.71 mg g(-1) (RSD = 3%) for carbon monoxide and 0.86 mg g(-1) (RSD = 3%) for carbon dioxide.
Related Concept Videos
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.