Sequential flow injection analysis of ammonium and nitrate using gas phase molecular absorption spectrometry
1Department of Chemistry, Institute for Advanced Studies in Basic Sciences, P.O. Box 45195-159, Gava Zang, Zanjan, Iran.
Abstract:
A flow injection method on the basis of gas phase molecular absorption is described for the sequential determination of ammonium and nitrate. Two hundred microliters of sample solution is injected into the flow line. For ammonium determination, the sample zone is directed to a line in which reacts with NaOH (13M) and produces ammonia. But for nitrate determination, the sample zone is passed through the on-line copperized zinc (Zn/Cu) reduction column and produces ammonium ion and in the follows ammonia. The produced ammonia in both cases is purged into the stream of N(2) carrier gas. The gaseous phase is separated from the liquid phase using a gas-liquid separator and then is swept into a flow through cell, which has been positioned in the cell compartment of an UV-Vis spectrophotometer. The absorbance of the gaseous phase is measured at 194nm. Under selected conditions for sequential analysis of ammonium and nitrate, linear relations were found between the peak heights of absorption signals and concentrations of ammonium (10-650mugml(-1)) and nitrate (20-800mugml(-1)). The limit of detections for ammonium and nitrate analysis were 8 and 10mugml(-1), respectively. The relative standard deviations of repeated measurements of 50mugml(-1) of ammonium and nitrate were 2.0, 2.9%, respectively. Maximum sampling rate was about 40 samples/h. The method was applied to the determination of ammonium in pharmaceutical products and the sequential determination of ammonium and nitrate in spiked water samples.
Related Concept Videos
Mass Spectrometry of Amines
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Mass Spectrometry: Amine Fragmentation
In amines, the number of nitrogen atoms affects the mass of the molecular ion, which is described by the nitrogen rule of mass spectrometry. This rule states that a compound containing a single or...
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Atomization Methods