Robust regression techniques A useful alternative for the detection of outlier data in chemical analysis
M Cruz Ortiz1, Luis A Sarabia, Ana Herrero
1Department of Chemistry, Faculty of Sciences, University of Burgos, Pza. Misael Bañuelos s/n, 09001 Burgos, Spain.
Abstract:
The validation of an analytical procedure means the evaluation of some performance criteria such as accuracy, sensitivity, linear range, capability of detection, selectivity, calibration curve, etc. This implies the use of different statistical methodologies, some of them related with statistical regression techniques, which may be robust or not. The presence of outlier data has a significant effect on the determination of sensitivity, linear range or capability of detection amongst others, when these figures of merit are evaluated with non-robust methodologies. In this paper some of the robust methods used for calibration in analytical chemistry are reviewed: the Huber M-estimator; the Andrews, Tukey and Welsh GM-estimators; the fuzzy estimators; the constrained M-estimators, CM; the least trimmed squares, LTS. The paper also shows that the mathematical properties of the least median squares (LMS) regression can be of great interest in the detection of outlier data in chemical analysis. A comparative analysis is made of the results obtained by applying these regression methods to synthetic and real data. There is also a review of some applications where this robust regression works in a suitable and simple way that proves very useful to secure an objective detection of outliers. The use of a robust regression is recommended in ISO 5725-5.
More Related Videos
06:50O-cresol Concentration Online Measurement Based On Near Infrared Spectroscopy Via Partial Least Square Regression
Published on: November 8, 2019
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry (UPLC-MS)
Published on: March 14, 2013
Related Concept Videos
Quantifying and Rejecting Outliers: The Grubbs Test
Outliers and Influential Points
Detection of Gross Error: The Q Test
Data Validation
Key parameters for method validation include:
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
What Are Outliers?
The z score is used to find outliers or unusual values. It should be noted that any values beyond -2 and +2 are...