Reliability of spinal range of motion in healthy boys using a skin-surface device
Eleftherios Kellis1, Grigoris Adamou, Gerasimos Tzilios
1Department of Physical Education and Sports and Sciences at Serres, Laboratory of Neuromuscular Control and Therapeutic Exercise, Aristotle University of Thessaloniki, Serres, Greece. ekellis@phed-sr.auth.gr
Objective:
The purpose of this study was to examine the interrater and intrarater reliability of spinal extension and flexion measurements using a skin-surface (Spinal Mouse; Idiag, Voletswil, Switzerland) instrument.
Methods:
Spinal curvature was measured during standing, full flexion, and full extension in 81 healthy children (10.6 +/- 1.7 years) by 3 raters on 2 separate occasions. Intraclass correlation coefficients (ICCs) and standard errors of measurement were used to examine between-day and interrater reliability for thoracic, lumbar, and hip range of motion.
Results:
The intrarater ICCs ranged from 0.61 to 0.96, and the interrater ICCs ranged from 0.70 to 0.93. The standard error of measurement ranged from 0.61 degrees to 13.18 degrees .
Conclusion:
Evaluation of spinal range of motion measures in children using the Spinal Mouse demonstrates fair to high reliability.