Related Experiment Video
Updated: Jun 28, 2026

Semi-Automated Planimetric Quantification of Dental Plaque Using an Intraoral Fluorescence Camera
Published on: January 27, 2023
[Determination of organic acids in dental plaque with high performance liquid chromatography]
Xiao-li Zou1, Hong-yan Zeng, Ci-ling Yan
1Dept. of Sanitary Technology, West China College of Public Health, Sichuan University, Chengdu 610041, China.
Objective:
To establish determination method of formic acid, lactic acid, acetic acid and succinic acid in dental plaque with high performance liquid chromatography (HPLC).
Methods:
After the samples were centrifuged, 2 microL supernatant was transferred to a 1 mL centrifuge tube and diluted in water, then was determined with HPLC. The mixture of phosphate buffer and methanol (97:3) as mobile phase throughout the experiment. The determination of organic acid was performed on Phenomenex C18 column and at their maximum absorption wave.
Results:
The linear ranges of formic acid, lactic acid, acetic acid and succinic acid were 0.110-500, 0.049-500, 0.047-500, 0.084-500 microg/mL. The detection limits were 0.110, 0.049, 0.047, 0.084 microg/mL. The relative standard derivation were 9.5%, 7.9%, 4.3%, 4.2%. The average recoveries of samples were 82%-112%, 82%-102.5%, 90%-115%, 80%-110%.
Conclusion:
The method was simple, quick and adapt for analysis of organic acid in dental plaque.