Related Experiment Video
Updated: Jun 27, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
A comparison of shielding calculation methods for multi-slice computed tomography (CT) systems
1KCARE, Medical Engineering and Physics, Faraday Building, King's College Hospital NHS Foundation Trust, London, UK. jonathan.cole@nhs.net
Abstract:
Currently in the UK, shielding calculations for computed tomography (CT) systems are based on the BIR-IPEM (British Institute of Radiology and Institute of Physics in Engineering in Medicine) working group publication from 2000. Concerns have been raised internationally regarding the accuracy of the dose plots on which this method depends and the effect that new scanner technologies may have. Additionally, more recent shielding methods have been proposed by the NCRP (National Council on Radiation Protection) from the USA. Thermoluminescent detectors (TLDs) were placed in three CT scanner rooms at different positions for several weeks before being processed. Patient workload and dose data (DLP: the dose length product and mAs: the tube current-time product) were collected for this period. Individual dose data were available for more than 95% of patients scanned and the remainder were estimated. The patient workload data were used to calculate expected scatter radiation for each TLD location by both the NCRP and BIR-IPEM methods. The results were then compared to the measured scattered radiation. Calculated scattered air kerma and the minimum required lead shielding were found to be frequently overestimated compared to the measured air kerma, on average almost five times the measured scattered air kerma.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...