Radiation dose values for various coronary calcium scoring protocols in dual-source CT
Paul Stolzmann1, Sebastian Leschka, Thomas Betschart
1University Hospital Zurich, Zurich, Switzerland.
Purpose:
The purpose of this study was to assess the radiation dose and associated image noise of previously suggested calcium scoring protocols using dual-source CT.
Methods:
One hundred consecutive patients underwent coronary calcium scoring using dual-source CT. Patients were randomly assigned to five different protocols: retrospective ECG-gating and tube current reduction to 4% outside the pulsing window at 120 (protocol A) and 100 kV (B), prospective ECG-triggering at 120 (C) and 100 kV (D), and prospective ECG-triggering at 100 kV with attenuation-based tube current modulation (E). Radiation dose parameters and image noise were determined and compared.
Results:
Protocol A resulted in an effective dose of 1.3 +/- 0.2 mSv, protocol B in 0.8 +/- 0.2 mSv, protocol C in 1.0 +/- 0.2 mSv, protocol D in 0.6 +/- 0.1 mSv, and protocol E in 0.7 +/- 0.1 mSv. Effective doses were significantly lower (P < 0.001) with 100 kV when compared to 120 kV protocols, and were significantly lower (P < 0.001) for prospective versus retrospective ECG-gating. No significant difference was found between protocol D and E. Significant negative correlations were found between the CTDI(vol) and heart rate for both retrospective ECG-gating protocols (protocol A: r = -0.98, P < 0.001; protocol B: r = -0.83, P < 0.001). The mean image noise was 29.0 +/- 6.7 HU, with no significant differences between the five protocols. The image noise was significantly correlated with the body weight (r = 0.21, P < 0.05) and BMI (r = 0.31, P < 0.01).
Conclusions:
Effective dose of calcium scoring using dual-source CT ranges from 0.6 to 1.3 mSv. Prospective triggering and lower tube voltage significantly reduces the radiation but yield similar image noise.
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...