Related Experiment Video
Updated: Jun 25, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Timing of the left ventricular electrogram and acute hemodynamic changes during implant of cardiac resynchronization
Berry M van Gelder1, Albert Meijer, Frank A Bracke
1Department of Cardiology, Catharina Hospital, Eindhoven, the Netherlands. carlgr@cze.nl
Study Objective:
To examine the relationship between timing of the left ventricular (LV) electrogram (EGM) and its acute hemodynamic effect on instantaneous change in LV pressure (LVdP/dt(MAX)).
Patients And Methods:
In 30 patients (mean = age 67 +/- 7.9 years) who underwent implant of cardiac resynchronization therapy systems, the right ventricular (RV) lead was implanted at the RV apex (n = 23) or RV septum (n = 7). The LV lead was placed in a posterior (n = 14) or posterolateral (n = 16) coronary sinus tributary. QRS duration, interval from Q wave to intrinsic deflection of the LV EGM (Q-LV), and interval between intrinsic deflection of RV EGM and LV EGM (RV-LV interval) were measured. The measurements were correlated with the hemodynamic effects of optimized biventricular (BiV) stimulation, using the Pearson correlation coefficient.
Results:
The mean LVdP/dt(MAX) at baseline was 734 +/- 180 mmHg/s, and increased to 905 +/- 165 mmHg/s during simultaneous BiV pacing, and to 933 +/- 172 mmHg/s after V-V interval optimization. The Pearson correlation coefficient R between QRS duration, the Q-LV interval, and the RV-LV interval at the respective LVdP/dt(MAX) was 0.291 (P = 0.66), 0.348 (P = 0.030), and 0.340 (P = 0.033).
Conclusions:
Similar significant correlations were observed between the acute hemodynamic effect of optimized BiV stimulation and the Q-LV and the RV-LV intervals. However, individual measurements showed an 80-ms cut-off for the Q-LV interval, beyond which the increase in LVdP/dt(MAX) was <10%.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrophysiology of Normal Cardiac Rhythm
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...