Reproducibility of the mfERG between instruments.
Wendy W Harrison1, Marcus A Bearse, Jason S Ng
1University of California Berkeley School of Optometry, 94720-2020, USA. wwh@berkeley.edu
Documenta Ophthalmologica. Advances in Ophthalmology
|March 27, 2009
Summary
Reproducibility of multifocal electroretinogram (mfERG) recordings is high when using Z-scores, even with different amplifiers or instruments. Converting mfERG data to Z-scores allows effective comparison across multicenter and longitudinal studies.
Area of Science:
- Ophthalmology
- Neuroscience
- Medical Instrumentation
Background:
- The multifocal electroretinogram (mfERG) is crucial for assessing retinal function.
- Variability in mfERG recordings across different instruments and amplifiers can hinder multicenter and longitudinal studies.
- Standardizing mfERG data acquisition and analysis is essential for reliable research.
Purpose of the Study:
- To evaluate the reproducibility and repeatability of mfERG recordings using different amplifiers and instrument versions.
- To establish a method for comparing and combining mfERG data from multiple recording instruments.
- To enhance the utility of mfERG in multicenter and longitudinal research.
Main Methods:
- mfERG recordings were made using two software versions (VERIS 4.3 and VERIS Pro 5.2) and three amplifiers.
- 73 subjects with normal vision and 26 with diabetes were tested across different recording setups.
- P1 implicit times and N1-P1 amplitudes were measured, and differences were analyzed as raw scores and Z-scores.
Main Results:
- Different amplifiers produced variations in mfERG implicit times, but amplitudes were comparable.
- mfERG recordings on different days and instruments showed more variability, particularly in amplitudes.
- Z-score conversion improved agreement between instruments, with a 95% confidence interval of approximately +/-1 Z-score for implicit times.
Conclusions:
- mfERG implicit times can differ significantly between amplifiers, even with identical settings.
- mfERG Z-scores demonstrate high reproducibility across different recording instruments.
- Converting mfERG data to Z-scores based on normative data is vital for comparing and combining data in multicenter and longitudinal studies.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
Uncertainty in Measurement: Accuracy and Precision
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.