Related Experiment Video
Updated: Jun 24, 2026

Preparation, Administration, and Assessment of In Vivo Tissue-Specific Cellular Uptake of Fluorescent Dye-Labeled Liposomes
Published on: July 30, 2020
Development and validation of a reverse phase liquid chromatography method for the quantification of rasagiline
Marcos Fernández1, Emilia Barcia, Sofía Negro
1Departamento de Farmacia y Tecnología Farmacéutica, Facultad de Farmacia, Universidad Complutense de Madrid, 28040 Madrid, Spain. marferna@udec.cl
Abstract:
In the present study, a reverse phase high performance liquid chromatographic method was developed and validated for the determination of rasagiline mesylate in biodegradable microspheres. Chromatographic separation was carried out on a RP-18 column using a mobile phase consisting of acetonitrile:water (5:95, v/v) adjusted at pH 3.1. Flow rate was 1.0 ml min(-1) and UV detection at 290 nm. Acyclovir was used as the internal standard. The calibration curve was linear over the range 0.5-20.0 microg ml(-1). R.S.D. for precision was <1.8%. Accuracy ranged between 99.01% and 102.55% with a R.S.D. lower than 1.3%. LOD and LOQ were 0.07 microg ml(-1) and 0.23 microg ml(-1), respectively. The method was simple, rapid, and easy to apply, making it very suitable for routine analysis of rasagiline mesylate in biodegradable PLGA microspheres. It could be also used with reliability for the determination of the drug in other pharmaceutical dosage forms.