Fabrication and Operation of an Oxygen Insert for Adherent Cellular Cultures
Published on: January 6, 2010
Performance of a MedArray silicone hollow fiber oxygenator
Nathan G LaFayette1, Rebecca E Schewe, J Patrick Montoya
1Department Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA.
Abstract:
A silicone hollow fiber oxygenator was evaluated to characterize gas transfer and biocompatibility. The device's fiber bundle was composed of MedArray's silicone hollow fibers with a 320 microm outside diameter, a 50 microm wall thickness, a surface area of 0.45 m, and a 0.49 void fraction. An in vitro gas exchange study was performed comparing the MedArray device (n = 9) with the Medtronic 0600 oxygenator (n = 6) using Association for the Advancement of Medical Instrumentation standards and blood flow rates of 0.5-1.75 L/min, and an oxygen to blood flow ratio of two. Biocompatibility and resistance studies were performed in vivo using a swine venovenous extracorporeal membrane oxygenation model (MedArray n = 5, Medtronic n = 5). Average O(2) transfer at 1 L/min was 86 ml/min/m in the MedArray device and 101.1 ml/min/m in the Medtronic device. At 0.5 L/min the MedArray and Medtronic device average resistance was 15.5 and 148.5 mm Hg/(L/min), respectively. Both devices had similar platelet consumption and hemolysis. Results indicate that the MedArray device has lower O(2) transfer efficiency, similar biocompatibility, and lower resistance than the Medtronic 0600 oxygenator. Optimization of the MedArray fiber bundle and housing design is necessary to improve O(2) transfer efficiency while maintaining lower device resistance than the Medtronic oxygenator.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: