Rejection of Fluorescence Background in Resonance and Spontaneous Raman Microspectroscopy
Published on: May 18, 2011
Method for automated background subtraction from Raman spectra containing known contaminants
Brooke D Beier1, Andrew J Berger
1The Institute of Optics, University of Rochester, Rochester, NY 14627, USA. beier@optics.rochester.edu
Abstract:
The use of Raman spectroscopy for biomedical applications requires overcoming the obstacle of the broad background that is also generated by biological samples. This background, which is often largely attributed to fluorescence, is frequently orders of magnitude greater than the Raman signal and needs to be removed in order to use Raman spectra in sample analysis. Several methods have been proposed for removing fluorescent signal, both instrumental and computational. Of the computational methods, polynomial fitting has become increasingly popular. Typically, a polynomial of approximately fifth order is used in the fitting. This method alone is not always capable of fitting some more tightly featured spectra that may be present in data, potentially coming from a contaminant in the sample itself or from the experimental design. If this signal is present in varying amounts, the polynomial background removal method can leave the residual spectra with non-uniform artifacts that hinder classification results. If a reference spectrum can be obtained for this interfering signal, however, it can be incorporated into the polynomial fit and removed separately. An automated method for the removal of broad and/or moderately featured background signal is described. In addition to simulations, the method has been applied to spectra from biofilms of Streptococcus mutans.
More Related Videos
07:37An Integrated Raman Spectroscopy and Mass Spectrometry Platform to Study Single-Cell Drug Uptake, Metabolism, and Effects
Published on: January 9, 2020
09:32Resolving Water, Proteins, and Lipids from In Vivo Confocal Raman Spectra of Stratum Corneum through a Chemometric Approach
Published on: September 26, 2019
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...