An unusual cause of asystole during a routine pacemaker generator change
Ilknur Can1, Venkat Tholakanahalli, Jian-Ming Li
1Fairview Medical Center, University of Minnesota, Minneapolis, Minnesota, USA.
Abstract:
A 74-year-old pacemaker-dependent male patient, who underwent a pacemaker generator change due to elective replacement indicator, had a 4-second pause during interrogation of the new pacemaker generator out of the device pocket due to a specific feature that was programmed on called"enhanced transtelephonic monitoring."