Related Experiment Video
Updated: Jun 21, 2026

Characterization of Nanocrystal Size Distribution using Raman Spectroscopy with a Multi-particle Phonon Confinement Model
Published on: August 22, 2015
Precise determination of Cr and Co in certified reference material of silicon nitride by neutron activation analysis
Tsutomu Miura1, Hideaki Matsue, Takayoshi Kuroiwa
1National Metrology Institute of Japan, AIST, Tsukuba, Ibaraki, Japan. t.miura@aist.go.jp
Abstract:
Neutron activation analysis with an internal standard correction was applied to the determination of Cr and Co in a ceramics certified reference material (NMIJ CRM 8004-a silicon nitride powder). Cesium was used as an internal standard to compensate for any inhomogeneity of the neutron flux through an irradiation capsule and to improve the repeatability of gamma-ray measurements. It was found that the linearity of the calibration curves of Cr and Co was improved by using an internal standard. The analytical results of Cr and Co in NMIJ CRM 8004-a were in good agreement with those obtained by ICP-OES, ICP-sector field mass spectrometry (ICP-SFMS), and isotope dilution/ICP-SFMS for Cr. The relative expanded uncertainties (k = 2) were 1.9% for Cr and 1.5% for Co. The uncertainties were comparable to those of atomic spectrometric methods.