[The omega connector--a module for jointed coupling of titanium total prostheses in the middle ear]
G Schmid1, U Steinhardt, W Heckmann
1HNO-Belegabteilung im Klinikum am Steinenberg, Reutlingen. g-fabbro@t-online.de
Background:
Hearing improvement after reconstruction of a defect ossicular chain depends on material, design and - crucially - coupling of the prosthesis. Coupling a total ossicular replacement prosthesis to the stapes footplate can be problematic and lead to prosthesis instability. In order to solve this problem, the Omega Connector was developed, a module allowing the middle ear surgeon to couple a titanium total prosthesis to the stapes footplate in a flexible, angle-variable manner.
Material And Method:
The Omega Connector is made of pure titanium and consists of three compenents: head, neck and base plate. The head allows a jointed coupling to the stem of a titanium total prosthesis. Positioned between the remnants of the stapes crura, the bas plate proves for a stable connection with the stapes footplate.
Results:
The Omega Connector was implanted during 14 revisions surgeries. In 10 of these surgeries a total ossicular replacement prosthesis was removed which was implanted at an earlier date and which was fixed and no longer functioning. In all 14 cases it was possible intraoperatively to position the Omega Connector correctly and to couple it to a titanium total prosthesis. The postoperative hearing gain was between 10 to 25 dB, with an average of 18 dB.
Conclusion:
For the first time, the Omega Connector offers to the surgeon the option to couple a titanium total prosthesis via a micro ball joint. Hearing results achieved so far are satisfying. The results confirm the advantages of the Omega Connector during implantation of a titanium total prosthesis. For a final evaluation, long-term studies have yet to be made.
More Related Videos
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
05:54Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
Related Concept Videos
The Auditory Ossicles
The aptly named stapes look very much like a stirrup. The three ossicles are unique to mammals, and each plays a role in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Anatomy of the Ear
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Bone as Supporting Connective Tissue
Bone Matrix
Bone, or osseous tissue, is a connective tissue that has a large amount of two different types of matrix material. The organic matrix is similar to the matrix material found in other connective tissues, including some amount of collagen and elastic fibers. This gives strength and flexibility to the tissue. The inorganic matrix consists of mineral salts— mostly calcium salts— that give the...
Structural Joints: Cartilaginous Joints
There are two types of cartilaginous joints:
Synchondrosis
A synchondrosis ("joined by cartilage") is a cartilaginous joint where bones are connected by hyaline cartilage. Synchondrosis may be temporary or...