Background:
- The RT3 triaxial accelerometer is a newer device intended to replace the TniTrac for activity monitoring.
- Validating new wearable technology against established methods is crucial for accurate physiological measurements.
Purpose of the Study:
- To validate the RT3 triaxial accelerometer against doubly labeled water (DLW) for measuring energy expenditure in free-living adults.
- To assess the accuracy of the RT3 in estimating total energy expenditure (TEE) and activity-related energy expenditure (AEE).
Main Methods:
- Thirty-six adults (18-56 years) participated in a 15-day study.
- Total energy expenditure (TEE) was measured using doubly labeled water (DLW) as the criterion method.
- The RT3 accelerometer was worn for 14 consecutive days to estimate TEE and AEE using its proprietary equation.
Main Results:
- The RT3 underestimated TEE by an average of 539 kJ (4%) compared to DLW.
- The RT3 underestimated AEE by an average of 485 kJ (15%) compared to DLW.
- RT3 measurements showed reasonable accuracy at the group level for AEE.
Conclusions:
- The RT3 triaxial accelerometer provides a relatively accurate group-level assessment of free-living activity-related energy expenditure.
- The device generally underestimates both total and activity-related energy expenditure when compared to DLW in adults.