Comparison of desaturation and resaturation response times between transmission and reflectance pulse oximeters
1Department of Anesthesiology and Pain Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Kangnam-Ku, Seoul, Korea.
Background:
In general, there is a response time between actual arterial hypoxemia and its detection by pulse oximeters. We compared the desaturation and resaturation response times between two types of pulse oximeters, transmission and reflectance pulse oximeters, to find out which oximeter has a more rapid response time.
Methods:
Thirty-three ASA 1 or 2 patients were enrolled in this study. A transmission pulse oximeter was placed on the index finger and a reflectance pulse oximeter was placed on the forehead and monitored simultaneously. After the induction of general anesthesia without pre-oxygenation, we waited until the oxygen saturation value of any of two pulse oximeters declined to 90%, and then mask ventilation was started with 100% oxygen. Oxygen saturation was recorded at an interval of 2 s during this time.
Results:
The desaturation response time of SpO(2) to 95% after apnea was 82.0 s (interquartile range: 67.0-98.5 s) vs. 94.0 s (interquartile range: 84.0-106.5 s) (P<0.001) and SpO(2) to 90% was 94.0 s (interquartile range: 75.5-109.5 s) vs. 100.0 s (interquartile range: 84.5-114.5 s) (P<0.001) in the reflectance and transmission oximeters, respectively. The resaturation response time from mask ventilation to 100% SpO(2) was 23.2+/-5.6 vs. 28.9+/-7.6 s (P<0.001) in the reflectance and transmission oximeters, respectively.
Conclusion:
In clinical situations in which rapid changes in oxygen saturation are expected, we recommend the forehead reflectance pulse oximeter because it responds more quickly in detecting oxygen desaturation and resaturation compared with the transmission pulse oximeter.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...