A biomechanical study comparing the compressive forces generated by a conventional 4.5 AO/ASIF cortical lag screw
B Lenehan1, J Street, B Murphy
1Department of Orthopaedic Surgery, Merlin Park Regional Hospital, Galway, Canada. blenenan@gmail.com
Abstract:
The aim of this study was to compare the interfragmentary compression generated across a simulated femoral fracture model by a conventional 4.5 mm AO/ASIF cortical lag screw with a differentially pitched cortical compression screw. A 45-degree osteotomy was made in a whole bone composite femoral shaft, this was internally fixed with either a conventional 4.5 mm AO/ASIF cortical lag screw or the differentially pitched cortical screw and the compressive force generated at the fracture site measured on an Instron 8874 Axial/Torsion Servohydraulic Testing System. The mean interfragmentary compression generated by the differentially pitched screw was 81.4% of that generated by the 4.5 mm AO/ASIF cortical lag screw. The 4.5 mm AO/ASIF cortical screw produces a steep rise in compression per turn of the screw. The screw based on the differential pitch design creates a more gradual increase to peak compression. The resistance to torque was greater for the AO screw than for the differential pitch screw. Maximal interfragmentary compression is achieved within 4 180 degrees turns after the head engages the near cortex for the 4.5 mm AO/ASIF screw but required 5 180 degrees turns for the differentially pitched screw. Interfragmentary compression is achievable in cortical bone using differential pitch technology. A differentially pitched screw offers obvious advantages over a conventional screw allowing independent placement of lag screw and neutralisation plate, without needing additional exposure of the fracture site, limiting the insult to local fracture biology. It is proposed as an adjunct to osteosynthesis in long bone fractures.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...
Normal Strain under Axial Loading
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Eccentric Axial Loading in a Plane of Symmetry