Metrological traceability of values for alpha-amylase catalytic concentration assigned to a commutable calibrator
Francesca Canalias1, Ester García, Maribel Sánchez
1Laboratori de Referència d'Enzimologia Clínica, Departament de Bioquímica i de Biologia Molecular, Unitat de Bioquímica de Medicina, Universitat Autònoma de Barcelona, Bellaterra, Spain. francesca.canalias@uab.cat
Background:
The International Standard ISO 18153 describes how to assure the metrological traceability of catalytic concentration values assigned to commercial calibration materials following the reference measurement system. We applied this approach to the standardization of alpha-amylase measurements.
Methods:
Traceable values of catalytic activity with measurement uncertainty were assigned to three commercial calibrator materials using alpha-amylase primary reference measurement procedure described by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). The metrological traceable to the SI units calibration was validated by measuring human serum and plasma samples using the primary reference procedure and two routine commercial measurement procedures using different substrates (4,6-ethylidene(G1)-4-nitrophenyl(G7)-alpha-(1-->4)-D-maltoheptaoside (EPS) and 2-chloro-4-nitrophenyl-alpha-D-maltotrioside (CNP)) and calibrated with the traceable materials. Previously the commutability of the commercial calibration materials was verified.
Results:
Procedures comparison showed that the primary reference procedure and routine procedures have the same analytical specificity. The commercial calibrators tested showed to be commutable and were used to recalibrate the routine measurement procedures. The agreement of procedures improves after recalibration with commutable calibrator materials.
Conclusions:
The implementation of a reference system for alpha-amylase measurements was demonstrated that assures the traceability of patients' results to SI units and to harmonize the results obtained by two different routine procedures.
More Related Videos
08:40Absolute Quantification of Aβ1-42 in CSF Using a Mass Spectrometric Reference Measurement Procedure
Published on: March 21, 2017
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Data Validation
Key parameters for method validation include:
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...