An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Photoelastic analysis of stress patterns from different implant-abutment interfaces
Sergio Rocha Bernardes1, Cleudmar Amaral de Araujo, Alfredo Julio Fernandes Neto
1Department of Occlusion, Fixed Prosthesis, and Dental Materials, Federal University of Uberlândia, School of Dentistry, Uberlândia, Brazil. sergiorb@rediffmail.com
Purpose:
To investigate the peri-implant stress fields generated from four different implant-abutment interfaces under axial loading applied at the center of the implant and several millimeters away from the implant center via photoelastic analysis.
Materials And Methods:
Similar unthreaded and cylindric implants and abutments were fabricated and embedded in photoelastic resin with four different implant-abutment interfaces: external hex, internal hex, internal taper (11.5 degrees), and solid connection to the abutment (one piece). The samples were submitted to vertical compressive loads; one was applied at the implant center (1.5 kg; centered load), and the other was applied 6.5 mm away from the center, 4.4 mm from the outside of the outer aspect of the implant (0.75 kg; off-centered load). The maximum shear stresses were determined and observed at 46 points around the implants under the centered load and at 61 points under the off-center load in the photoelastic models. Graphics describing the maximum shear stress (y-axis) and the analyzed points (x-axis) were obtained, and areas under the curves were calculated.
Results:
The centered loading (all points) resulted in small differences. The lowest amounts of stress were observed for the internal-taper implants, and values were minimally greater (0.4% to 3.3%) for the other implants. No statistically significant differences were found between groups for the centered load in any area. Under an off-center load, the internal-hex implants presented the least stress (all points). For off-center loading, the internal-hexagon implants differed significantly from the external-hex and one-piece implants and displayed the lowest stress levels.
Conclusion:
Under an off-center load, the internal-hex interfaces presented the lowest stress concentrations, internal-taper interfaces presented intermediate results, and one-piece and external-hex implants resulted in high stress levels. Centralized axial loads produced similar results.
More Related Videos
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes.
Transformation of Plane Stress
Stress Concentrations
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and opposite to those on the...