TG-43 U1 based dosimetric characterization of model 67-6520 Cs-137 brachytherapy source
Ali S Meigooni1, Clarissa Wright, Rafiq A Koona
1Department of Radiation Medicine, North Shore University Hospital, 300 Community Drive, Manhasset, New York 11030, USA. alimeig@gmail.com
Purpose:
Brachytherapy treatment has been a cornerstone for management of various cancer sites, particularly for the treatment of gynecological malignancies. In low dose rate brachytherapy treatments, 137Cs sources have been used for several decades. A new 137Cs source design has been introduced (model 67-6520, source B3-561) by Isotope Products Laboratories (IPL) for clinical application. The goal of the present work is to implement the TG-43 U1 protocol in the characterization of the aforementioned 137Cs source.
Methods:
The dosimetric characteristics of the IPL 137Cs source are measured using LiF thermoluminescent dosimeters in a Solid Water phantom material and calculated using Monte Carlo simulations with the GEANT4 code in Solid Water and liquid water. The dose rate constant, radial dose function, and two-dimensional anisotropy function of this source model were obtained following the TG-43 U1 recommendations. In addition, the primary and scatter dose separation (PSS) formalism that could be used in convolution/superposition methods to calculate dose distributions around brachytherapy sources in heterogeneous media was studied.
Results:
The measured and calculated dose rate constants of the IPL 137Cs source in Solid Water were found to be 0.930 (+/-7.3%) and 0.928 (+/-2.6%) cGy h(-1) U(-1), respectively. The agreement between these two methods was within our experimental uncertainties. The Monte Carlo calculated value in liquid water of the dose rate constant was null set=0.948 (+/-2.6%) cGy h(-1) U(-1). Similarly, the agreement between measured and calculated radial dose functions and the anisotropy functions was found to be within +/-5%. In addition, the tabulated data that are required to characterize the source using the PSS formalism were derived.
Conclusions:
In this article the complete dosimetry of the newly designed 137Cs IPL source following the AAPM TG-43 U1 dosimetric protocol and the PSS formalism is provided.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...