Contrast material administration protocols for 64-MDCT angiography: altering volume and rate and use of a saline
Courtney A Coursey1, Rendon C Nelson, Paul W Weber
1Department of Radiology, Duke University Medical Center, Durham, NC 27710, USA.
Objective:
The purpose of our study was to evaluate the effect of varying volumes and rates of contrast material, use of a saline chaser, and cardiac output on aortic enhancement characteristics in MDCT angiography (MDCTA) using a physiologic phantom.
Materials And Methods:
Volumes of 75, 100, and 125 mL of iopamidol, 370 mg I/mL, were administered at rates of 4, 6, and 8 mL/s. The effect of a saline chaser (50 mL of normal saline, 8 mL/s) was evaluated for each volume and rate combination. Normal, reduced (33% and 50%), and increased (25%) cardiac outputs were simulated. Peak aortic enhancement and duration of peak aortic enhancement were recorded. Analysis of variance models were run with these effects, and the estimated mean levels for the sets of factor combinations were determined.
Results:
Lowering the volume of contrast material resulted in reduced peak enhancement (example, -56.2 HU [p < 0.0001] with 75 vs 125 mL) and reduced duration of 75% peak enhancement (example, -9.0 seconds [p < 0.0001] with 75 vs 125 mL). Increasing the rate resulted in increased peak enhancement (example, 104.5 HU [p < 0.0001] with a rate of 8 vs 4 mL/s) and decreased duration of 75% peak enhancement (example, -13.0 seconds [p < 0.001]). Use of a saline chaser resulted in increased peak enhancement, and this increase was inversely proportional to contrast material volume. Peak enhancement increased when reduced cardiac output was simulated. Peak enhancement decreased when increased cardiac output was simulated.
Conclusion:
Reducing contrast material volume from 125 to 75 mL, increasing the rate to 6 or 8 mL/s, and use of a saline chaser result in an aortic enhancement profile that better matches the approximately 5-second imaging window possible with 64-MDCTA of the abdomen and pelvis. Even smaller volumes of contrast material may be adequate in patients with reduced cardiac output.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...