A randomized double-blind crossover comparison of four rate-responsive pacing modes
N Sulke1, J Chambers, A Dritsas
1Department of Cardiology, Guy's Hospital, London, England.
Journal of the American College of Cardiology
|March 1, 1991
Summary
Dual-sensor rate-responsive pacing (DDDR) offers superior objective and subjective benefits compared to single-sensor modes (VVIR, DDIR, DDD) for patients with pacemakers. DDDR pacing enhances well-being and exercise capacity, making it the preferred choice.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Rate-responsive pacemakers adjust heart rate based on physiological demand.
- Modern pacemakers offer various pacing modes with different sensor functionalities.
- Optimizing pacemaker mode selection is crucial for patient quality of life and physiological response.
Purpose of the Study:
- To compare subjective and objective outcomes of four modern rate-responsive pacemaker modes.
- To evaluate patient preference and physiological performance across VVIR, DDIR, DDD, and DDDR pacing modes.
- To determine the superiority of dual-sensor pacing over single-sensor modes.
Main Methods:
- Double-blind crossover study involving 22 patients with DDDR pacemakers.
- Randomized programming to VVIR, DDIR, DDD, or DDDR modes for 4 weeks each.
- Subjective assessments (well-being, symptoms) and objective measures (exercise capacity, cardiac output, regurgitation) were performed.
Main Results:
- DDDR mode was preferred by 59% of patients; VVIR was least acceptable in 73%.
- VVIR pacing showed significantly worse well-being, exercise capacity, and functional status compared to dual-rate-responsive modes.
- DDDR pacing demonstrated longer exercise treadmill times and improved cardiac output at rest.
- Heart rate response to mental stress was suboptimal in VVIR and DDIR modes.
- Tricuspid regurgitation was significantly greater in VVIR mode than in dual modes.
Conclusions:
- Dual-sensor DDDR pacing is objectively and subjectively superior to single-sensor VVIR, DDIR, and DDD pacing.
- DDDR pacing improves patient well-being, exercise capacity, and hemodynamic function.
- Subjective benefits of dual-chamber rate-augmented pacing can be predicted echocardiographically.