Determination of platinum traces contamination by graphite furnace atomic absorption spectrometry after
M Chappuy1, E Caudron, A Bellanger
1Analytical Development Laboratory, Pharmaceutical Establishment of Paris Hospitals, 7 rue du Fer à Moulin, 75005 Paris, France.
Abstract:
A simple and sensitive method is described for the determination of platinum surface contamination originating from cisplatin, carboplatin and oxaliplatin. Following extraction from swabs and preconcentration with the cloud point extraction (CPE) method, detection was by graphite furnace atomic absorption spectrometry (GFAAS). After desorption of platinum compounds from the swab, CPE involved on preconcentration of platinum in aqueous solution with diethyldithiocarbamate (DDTC) as chelating agent and Triton X-114 as extraction medium. DDTC is not only a chelating agent, but may also be a good candidate for the inactivation of platinum compounds. DDTC is recommended by the Word Health Organization (WHO) for the destruction of platinum-based anticancer drugs. The main factors affecting CPE efficiency, pH of the sample solution, concentrations of DDTC and Triton X-114, equilibration temperature and incubation time, were evaluated in order to enhance sensitivity of the method. The desorption of platinum compounds from the swab was investigated in parallel. Since platinum is bound to DDTC, it must exchange with copper in order to enhance platinum atomizing by GFAAS. A preconcentration factor of 29 was obtained for 10 mL of a platinum solution at 10 microg mL(-1). In optimal conditions, the limit of detection was 0.2 ng mL(-1), corresponding to 2.0 ng of platinum metal on the swab. Absorbance was linear between 0.7 and 15 ng mL(-1). The proposed method was applied for the determination of surface contamination by platinum compounds with correct results.
More Related Videos
12:12On the Preparation and Testing of Fuel Cell Catalysts Using the Thin Film Rotating Disk Electrode Method
Published on: March 16, 2018
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Absorption Spectroscopy: Atomization Methods
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Washing, Drying, and Ignition of Precipitates
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...