Related Experiment Video
Updated: Jun 17, 2026

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Remote magnetic catheter navigation for cavotricuspid isthmus ablation in patients with common-type atrial flutter
Dirk Vollmann1, Lars Lüthje, Joachim Seegers
1Division of Cardiology and Pneumology, Heart Center, University of Göttingen, Göttingen, Germany. d.vollmann@med.uni-goettingen.de
Background:
Conventional catheter ablation of cavotricuspid isthmus (CTI)-dependent atrial flutter is a widely applied standard therapy. Remote magnetic catheter navigation (RMN) may provide benefits for different ablation procedures, but its efficacy for CTI ablation has not been evaluated in a randomized, controlled trial.
Methods And Results:
Ninety patients undergoing de novo ablation of atrial flutter were randomly assigned to conventional manual (n=45) or RMN-guided (n=45) CTI ablation with an 8-mm-tip catheter. Complete bidirectional isthmus block was achieved in 84% (RMN) and 91% (conventional catheter ablation) of the cases (P=0.52). RMN was associated with shorter fluoroscopy time (median, 10.6 minutes; interquartile range [IQR], 7.6 to 19.9, versus 15.0 minutes; IQR, 11.5 to 23.1; P=0.043) but longer total radiofrequency application (17.1 minutes; IQR, 8.6 to 25, versus 7.5 minutes; IQR, 3.6 to 10.9; P<0.0001), ablation time (55 minutes; IQR, 28 to 76, versus 17 minutes; IQR, 7 to 31; P<0.0001), and procedure duration (114+/-35 versus 77+/-24 minutes, P<0.0001). Procedure duration in the RMN group did not decrease significantly with case experience. Long-term procedure success, defined as achievement of complete CTI block and freedom from atrial flutter recurrence during 6 months of follow-up, was lower in the RMN group (73% versus 89%, P=0.063). Right atrial angiography after ablation revealed no significant differences between groups in terms of right atrial diameter or CTI length, morphology, and angulation. Furthermore, none of these parameters was predictive for difficult (ablation time >20 minutes) or unsuccessful ablation.
Conclusions:
RMN-guided CTI ablation is associated with reduced radiation exposure but prolonged ablation and procedure times as compared with conventional catheter navigation. Our findings suggest that ablation lesions produced with an RMN-guided 8-mm catheter are less effective irrespective of CTI anatomy.
Trial Registration:
clinicaltrials.gov Identifier: NCT00560872.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization III: Left Heart Catheterization