Related Experiment Video
Updated: Jun 17, 2026

Multi-analyte Biochip (MAB) Based on All-solid-state Ion-selective Electrodes (ASSISE) for Physiological Research
Published on: April 18, 2013
Development of a micro-potentiometric sensor for the microchip analysis of alkali ions
Adelina Smirnova1, Kazuma Mawatari, Hiroko Takahashi
1Microchemistry Group, Kanagawa Academy of Science and Technology, 3-2-1 Sakado, Takatsu, Kawasaki 213-0012, Japan.
Abstract:
This paper reports on the development of a micro-potentiometric sensor based on external microelectrodes introduced into a microchip. We miniaturized reference and ion-selective electrodes (ISEs) and embedded them into a plastic (PDMS) microchip; the miniaturization of ISE was attained by using a monolithic capillary-based membrane. This sensor was applied to the detection of alkali ions (Na+, K+ and NH4+) in a microflow on the microg/L level.
Related Concept Videos
Potentiometry: Membrane Electrodes
Potentiometry: Overview
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...