Calibration coefficient of reference brachytherapy ionization chamber using analytical and Monte Carlo methods
Sudhir Kumar1, P Srinivasan, S D Sharma
1Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai 400094, India.
A new graphite ionization chamber was developed for measuring high dose rate Iridium-192 brachytherapy sources. Both analytical and Monte Carlo methods validated its calibration, ensuring accurate radiation dosimetry.
Area of Science:
- Medical Physics
- Radiation Dosimetry
- Nuclear Engineering
Background:
- Accurate measurement of high dose rate (HDR) (192)Ir brachytherapy sources is crucial for effective cancer treatment.
- Reference dosimeters are essential for calibrating radiation sources and ensuring treatment accuracy.
- Existing methods for calibrating HDR sources require precise and reliable dosimetric tools.
Purpose of the Study:
- To design and fabricate a cylindrical graphite ionization chamber for use as a reference dosimeter.
- To determine the air kerma calibration coefficient (N(K)) of the fabricated ionization chamber using analytical and Monte Carlo methods.
- To experimentally validate the performance of the ionization chamber by measuring the reference air kerma rate (RAKR) of HDR (192)Ir sources.
Main Methods:
- A 1002.4 cm(3) cylindrical graphite ionization chamber was designed and fabricated at Bhabha Atomic Research Centre (BARC).
- Analytical estimation of N(K) using Burlin general cavity theory, calculating coefficients for each spectral line of an HDR (192)Ir source.
- Monte Carlo simulation using the Monte Carlo N-particle code to model the measurement setup, HDR (192)Ir source, and surrounding materials.
Main Results:
- The Monte Carlo calculated N(K) showed agreement within 1.77% compared to the analytical method.
- Experimentally determined RAKR of HDR (192)Ir sources using the reference ionization chamber agreed within 1.43% with vendor-quoted values.
- The developed ionization chamber serves as a reliable reference dosimeter for HDR (192)Ir brachytherapy sources.
Conclusions:
- The designed graphite ionization chamber is suitable for accurate dosimetry of HDR (192)Ir brachytherapy sources.
- Both analytical and Monte Carlo methods provide consistent results for the air kerma calibration coefficient.
- The experimental validation confirms the chamber's capability to measure reference air kerma rates accurately.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Correlation Coefficient
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...