Biomechanical comparison of contemporary clavicle fixation devices
Timothy Renfree1, Bryan Conrad, Thomas Wright
1Department of Orthopaedics and Rehabilitation, University of Florida, Gainesville, FL 32607, USA.
Purpose:
Because recent studies have shown that malunited or nonunited clavicle fractures treated nonsurgically have poor outcomes, early fixation of certain clavicle shaft fractures using contemporary implants has become more common. Little is known about the physiologic loading of these implants. This study was designed to observe the biomechanical behavior and strength of implants used for fixation of the clavicle shaft.
Methods:
We used synthetic clavicle bones that were produced in 2 pieces to replicate the conditions of a transverse fracture. Ten specimens per group were tested using 3 clavicle fixation constructs: precontoured clavicle plates with unicortical locking, precontoured clavicle plates with bicortical nonlocking screws, and intramedullary Rockwood clavicle pins. Two loading conditions were used: cantilever bending and 3-point bending.
Results:
For cantilever bending, both plate groups failed by fracture at the most medial screw hole. The pin group was unable to resist the small torque associated with off-axis loading. The pin group failed at a lower maximum load, which occurred with a greater maximum displacement. The pin group was also less stiff than the plate group. In the 3-point bending test, the pin group was significantly more flexible and achieved higher failure loads than the plate groups. There were no significant differences between the unicortical and bicortical screws for any of the measured variables with either loading condition.
Conclusions:
Both locking and nonlocking constructs appear to provide similar rigid fixation under the tested mechanical conditions. The intramedullary pin can provide high resistance to failure loads in situations when rigidity and rotational stiffness are not required. Intramedullary pin fixation appears to be inadequate where rotational stiffness is required.
More Related Videos
05:45Author Spotlight: Exploring Regeneration in Axolotls Through Insights in Cellular and Molecular Mechanisms, Bone Healing, and Implications for Human Therapies
Published on: April 12, 2024
06:51Methods for In Vivo Biomechanical Testing on Brachial Plexus in Neonatal Piglets
Published on: December 19, 2019
Related Concept Videos
Carbon-dioxide Fixation
Contemporary Psychology
Biopsychology
Biopsychology, also known as biological psychology or behavioral neuroscience, focuses on the biological underpinnings of behavior and mental processes. It...
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons