Related Experiment Video
Updated: Jun 16, 2026

Development of a Nose-only Inhalation Toxicity Test Chamber That Provides Four Exposure Concentrations of Nano-sized Particles
Published on: March 18, 2019
Development of simulated air filters for gamma-ray spectrometry proficiency testing
A Ceccatelli1, P De Felice, A Fazio
1ENEA, Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (INMRI), Centro Ricerche Casaccia, Rome, Italy.
Abstract:
A set of 60 spiked simulated air filter sources was developed for the 2004-2005 proficiency test of the Italian radioactivity surveillance network. Each simulated filter source was prepared by reproducible volumetric dispensing of 19 drops of a mixed radioactive solution containing Co-57, Cs-134, Cs-137, Co-60 and Eu-152. The spiking pattern was investigated and optimized by numerical simulations. The filter sources were checked by double counting on HPGe and NaI(Tl) detectors. The standard deviation (SD) of activity values was 1.5%.