Statement Of Problem:
Dentistry is gravitating toward the increased use of electric handpieces. The dental professional should have sufficient evidence to validate the switch from an air-turbine handpiece to an electric handpiece. However, there is little research quantifying the cutting efficiency of electric and air-turbine handpieces. Studies that do quantify cutting efficiency typically do so with only a single material.
Purpose:
The purpose of this study was to compare the cutting efficiency of an electric handpiece and an air-turbine handpiece, using various materials commonly used in dentistry.
Material And Methods:
Seven materials: Macor (machinable glass ceramic), silver amalgam, aluminum oxide, zirconium oxide, high noble metal alloy, noble metal alloy, and base metal alloy, were each cut with a bur 220 times; 110 times with an electric handpiece, and 110 times with an air-turbine handpiece. The weight difference of the material was calculated by subtracting the weight of the material after a cut from the weight of the material before the cut. The cutting efficiency was calculated by dividing the weight difference by the duration of the cut (g/s). Data were analyzed by a 2-way analysis of variance followed by Tukey's Honestly Significant Difference (HSD) test (alpha=.05).
Results:
The electric handpiece cut more efficiently than the air-turbine handpiece (F=3098.9, P<.001). In particular, the high noble metal alloy, silver amalgam, and Macor were cut more efficiently with the electric handpiece (0.0383 +/-0.0002 g/s, 0.0260 +/-0.0002 g/s, and 0.0122 +/-0.0002 g/s, respectively) than with the air-turbine handpiece (0.0125 +/-0.0002 g/s, 0.0142 +/-0.0002 g/s, and 0.008 +/-0.0002 g/s, respectively).
Conclusions:
The electric handpiece is more efficient at cutting various materials used in dentistry, especially machinable glass ceramic, silver amalgam, and high noble alloy, than the air-turbine handpiece.
More Related Videos
Related Concept Videos
Wind Turbine Machine Models
Induction machines interact through the rotating magnetic field generated by the stator and the rotor. The key parameter is slip, which is the difference between synchronous speed and rotor speed relative to synchronous speed. Slip is...
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
Conservation of Energy in Control Volume
For steady flow systems, the time derivative of the stored energy becomes zero since there is no energy accumulation within the control volume. This simplifies the energy equation to:
Otto and Diesel Cycle