Mycobacterium chelonae valve endocarditis resulting from contaminated biological prostheses
T M V Strabelli1, Rinaldo Focaccia Siciliano, Jussara Bianchi Castelli
1Heart Institute InCor, Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, Brazil.
Rapid-growing mycobacteria caused biological prosthetic valve endocarditis in Brazil, linked to manufacturing. Ziehl-Neelsen staining of prosthetic valve endocarditis is crucial, especially with negative blood cultures.
Area of Science:
- Infectious Diseases
- Cardiovascular Surgery
- Microbiology
Background:
- Biological prosthetic valve (BPV) endocarditis can be challenging to diagnose, particularly when causative agents are not readily identified through standard blood cultures.
- A potential link between prosthetic manufacturing processes and microbial contamination of BPVs warrants investigation.
Purpose of the Study:
- To describe a series of biological prosthetic valve endocarditis cases in Brazil potentially linked to the prosthetic manufacturing process.
- To identify the causative agent and evaluate diagnostic methods for culture-negative BPV endocarditis.
Main Methods:
- Review of clinical and microbiological data from 39 patients undergoing BPV replacement for suspected endocarditis between 1999 and 2008.
- Histological examination of BPV explants using Ziehl-Neelsen staining for acid-fast bacilli (AFB).
- Molecular identification of AFB using 16S-23S internal transcribed sequence (ITS) analysis on DNA extracted from paraffin-embedded samples.
Main Results:
- 46% (18/39) of BPV explants were positive for AFB, with 13 cases confirmed as endocarditis and 5 as colonization.
- AFB were detected in 5 out of 6 "sterile" non-implanted BPVs from the same manufacturer.
- Molecular analysis identified Mycobacterium chelonae as the causative agent.
Conclusions:
- Rapid-growing mycobacteria, specifically Mycobacterium chelonae, should be suspected in cases of BPV endocarditis, particularly those associated with prosthetic manufacturing.
- The Ziehl-Neelsen stain on histological sections is a critical diagnostic tool for BPV endocarditis, especially when blood cultures are negative.
- Early or late BPV endocarditis may be linked to contaminated prostheses, highlighting the importance of stringent manufacturing and sterilization processes.